MAN 01/EN/03 CRONO PAR 50 06/13
FrEE
BUTTONS CONFIGURING THE SETTINGS IN THE OFF CONDITION DISPLAY
PUMP SET TO OFF
• Pump set to OFF
• Set bolus dose volume (from 0.00 to 2.00 ml)
PROG
• Setting the interval between the bolus doses
(from “no,Lt” to 24 hours)
PROG
• Setting the ow rate F1 (0.05 to 5.00)
PROG
ml/h
• Setting the ow rate F2 (0.00 to 5.00 - oFF - disabled)
PROG
ml/h
• Setting the ow rate F3 (0.00 to 5.00 - oFF - disabled)
PROG
ml/h
• Setting the partial volume (from 1 to 50 ml)
(see table SETTING THE PARTIAL VOLUME)
PROG
• Pump set to OFF
/
• Decrease/increase stored value
SECTION 10
BUTTONS SETTING THE PARTIAL VOLUME DISPLAY
PUMP SET TO OFF
• Setting the partial volume
PROG
• Decrease/Increase of partial volume
PROG
• Positioning pusher to partial volume
• Pump set to OFF
/