Cane Crono Guía de usuario Pagina 41

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 114
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 40
41
MAN 01/EN/03 CRONO PAR 50 06/13
FrEE
FrEE
SETTING THE SECOND AND THIRD FLOW RATES (F2 AND F3)
The ow rate can be set from 0.00 to 5.00 ml/h as follows:
from 0.00 ml/h to 1.00 ml/h in increments of 0.01 ml/h;
from 1.00 ml/h to 3.00 ml/h in increments of 0.02 ml/h;
from 3.00 ml/h to 5.00 ml/h in increments of 0.05 ml/h.
The second and third ow rates are set in the same manner
as the rst.
After you have set the third ow, press the
button before
the OFF or StoP indication is displayed (while the F3 hourly
ows value is still ashing) to continue to the setting of the
next parameter: PARTIAL VOLUME.
If you keep either the
or button pressed, it is possible
to change the ow rate quickly.
If the ow rate is programmed to 0.00 ml/h, the device
will not administer the drug via ow, but only through the
administration of bolus doses.
it is also possible to set the oFF condition for ows F2 and/
or F3, so that the ows will not be displayed, and therefore
will not be selectable, when the pump is ON.
NOTES
SECTION 11
PROG
PROG
PROG
ml/h
PROG
ml/h
PROG
ml/h
Vista de pagina 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 113 114

Comentarios a estos manuales

Sin comentarios